MEDTRONIC PUERTO RICO OPERATIONS CO. Vanta; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
| Model Number |
977006 |
| Medical Device Problem Codes |
High impedance (1291); Low impedance (2285)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with an implantable neurostimulator system fell on the left buttock/hip a few days before the initial report and reported continued pain relief and a difference in normal pain.The patient reported discomfort, soreness, or pinching and was alive with no injury.A postoperative evaluation was performed, and stimulation thresholds were set for all groups.The patient was instructed to gradually adjust stimulation as needed for comfort.Impedance testing identified high impedance greater than 40,000 on contact 2.Contacts 4 and 13 were within normal limits but reflected a possible short; contacts 2, 4, and 13 were not used for programming.No changes were made other than setting stimulation thresholds.Imaging was not planned unless the patient developed increased pain.The manufacturer representative (rep) indicated they would send any further information as they become aware.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|