(b)(4).Date sent: 9/17/2026.D4: batch # 489c82.Additional information was requested, and the following was obtained:
1.Was the firing sequence started but not completed? yes.2.If yes, did the device deliver any staples? no.3.What was the appearance of the deployed staples (b-formed, malformed, goal post/legs straight)? na.4.Were there staples missing from the staple line? no.5.If yes, was the staple line complete? na.6.Did the device cut? no.7.If yes, was the cut line complete? na.8.If yes, was there any issue with the cut line? na.9.Was there any difficulty opening the device jaws? no.10.If yes, what steps were taken to open the jaws? na.11.If the jaws could not be opened, how was the device removed? na.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst60g reload was received partially fired 1/4 and with an opened package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 489c82, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number of 759d04/23gst60g, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|