| Catalog Number |
515070 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following information was provided by the initial reporter:
customer advised that there was a leak of azacitdine when being delivered via subcutaneous injection from phaseal optima to needle connection.Advised that they put connector into injector then added needle to connector.Provided education of the order in which it needs to be connected- needle into connector then connector into injector
additional information:
the leak occurred after assembling and when the therapy was being administered.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|