| Catalog Number |
PW100 |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pure wick urine collection system were used with the replacement canister.However, after overnight use, the canister remained empty, and the lack of suction resulted in urine leakage.The customer was transferred to the medical information department, where a water suction test was performed (removing the catheter and placing the bare tubing into a cup of water while the machine was running).Despite the motor operating and the power indicator light being on, the device failed to draw any water, confirmed a suction failure.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|