• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Diabetes Care Ltd FreeStyle Libre 3; Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Diabetes Care Ltd FreeStyle Libre 3; Continuous Glucose Monitoring System Back to Search Results
Model Number 72079-01
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported their adc device is too hot to hold.There was no report of fire, smoke, explosion, or shock.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
Reader (b)(6) has been returned and investigated.Visual inspection has been performed and the serial number was observed to be rubbed off.No evidence of the reader being too hot to hold was observed.A built in reader test was performed and results were within specification.De-cased the returned reader and performed visual inspection of the printed circuit board assembly (pcba), no signs of damage or burn marks were observed.Liquid contamination was observed on the pcba.The returned battery was measured and results were within specification.Using a thermal camera the reader was monitored and the temperature was observed to be within specification.Power consumption was not able to be performed due to liquid contamination.The observed liquid contamination observed is the result of foreign ingress introduced to the pcba while in the hands of the customer.This contamination is not an indication of a product failure and is the result of misuse.Furthermore, the lack of hot spots on the pcba through the thermal camera is an indication the electronics of the reader were not producing excessive heat and thus is not the cause of the complaint.Therefore, this issue not confirmed to use.This complaint has been reassessed as reportable upon retrospective review associated with a capa and is being submitted with all available information.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FreeStyle Libre 3
Common Device Name
Continuous Glucose Monitoring System
Manufacturer (Section D)
Abbott Diabetes Care Ltd
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
melissa pieplow
1360 south loop road
alameda, CA 94502--700
+1(510)384-5975
MDR Report Key50803018
Report Number2954323-2026-420934
Device Sequence Number10589719
Product Code QBJ
UDI-Device Identifier00357599820008
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K223435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number72079-01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/27/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-