The account reported finding rust on instruments, with about 20 looking "rusty" in some areas or having some kind of red fuzz on it.A photo and a sample were received for investigation; three photos and three opened kits were reviewed, and small spots were observed on the scissors, with no rust seen on the other blade, between the hinges, on the forceps component, or other area of the kit.Small spots of what appeared to be rust were observed on scissors item 105916 scs litt fg 4.5, confirming the reported issue.The vendor reviewed the returned samples and determined that the root cause is result of a burn mark that was made due to the steel meeting the polishing brush, resulting in "yellow tarnish" and not rust; no additional action was taken by the vendor at this time, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|