| Model Number |
CPR4 HEAD |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportedly, head installed as a replacement for ra2211068, which has cracks on two of the three corners with mounting points.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the provided pictures, it was confirmed that the telemetry head has slight cracks but is still functioning correctly, as it was used to replace a broken telemetry head.The most probable hypothesis is that a fall on a hard surface caused the cracks on the plastic parts.This case is retained and utilized for trend purposes.
|
| |
|