| Model Number |
MENISCAL CINCH¿ II |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 3rd september 2026, it was reported by an arthrex's employee via an email that for 3 units of ar-4501, meniscal cinch¿ ii:
the first unit of ar-4501's anchor dropped out when the suture was tightened.A 2nd ar-4501 was changed to but both anchors dropped out too.A 3rd ar-4501 was used but the second anchor dropped out again.This was detected during a knee meniscus repair surgery on (b)(6) 2026.The procedure was completed successfully by using a 4th ar-4501.There were no patient harm and no secondary procedure needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|