The customer states they used only 3 of the syringes out of the entire case and 2 of them fell apart, states the needle is unstable, and reported that during administration the entire hub shot off from the syringe lost half of solution.Medline received 1cs, syr101272 standard hypodermic syringe with needle in new condition, randomly selected 50 samples to perform a visual inspection and functional testing by inserting each syringe into a training pad, and tested for leakage after fluids were drawn from a training vial and then flushed.The complaint was not confirmed: the visual inspection revealed no manufacturing or any other defects, there were no indications of any of the needles detaching from their hub, all needles and syringes remained attached at all times, and no leakage was seen from the needle, hub, or plunger/barrel.The cause was documented as unconfirmed defect / not confirmed with received sample, the samples worked as intended, and there were no reports of patient injury or indications of nonconforming product based on the investigation details.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|