• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Resolution 360 Clip; Hemostatic Metal Clip For The Gi Tract

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Resolution 360 Clip; Hemostatic Metal Clip For The Gi Tract Back to Search Results
Model Number M00521231
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a resolution 360 clip was used during a procedure.During the procedure, when the clip was deployed it failed to release from the catheter.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the procedure details to date, despite good faith efforts.
 
Additional Manufacturer Narrative
Block h6: imrdf device code a15 captures the reportable event of clip failure to release from catheter.Block b3: approximated based on the date the manufacturer became aware of the event.Block h11: investigation results the resolution 360 clip was not returned for analysis; therefore, a technical analysis could not be performed, and for this reason and due to the lack of evidence is unable to confirm the reported event.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Based on the event description and information provided by the customer there is not enough evidence to determine whether the issue was induced due to the physicians technique during the procedure or was related to a device malfunction.Additionally, based on the most relevant information for the complaint including product record review results, the device met all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process, the definitive cause of the reported issue cannot be established due to lack of evidence.Without proper evaluation of the device, it remains unknown the causes that contributed to the event.Based on the thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Resolution 360 Clip
Common Device Name
Hemostatic Metal Clip For The Gi Tract
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50803374
Report Number3005099803-2026-03828
Device Sequence Number10611023
Product Code PKL
UDI-Device Identifier08714729875635
UDI-Public(01)08714729875635(17)290515(10)39352684
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K222503
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00521231
Device Catalogue NumberM00521231
Device Lot Number0039352684
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/17/2026
Date Device Manufactured05/16/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-