The reported issue was leaking; per medwatch, it was discovered in the cath lab that particulates from a medication core leaked into the medication of a multi-dose vial.A sample(s) was not received for investigation, and the complaint could only be confirmed by further investigating the sample or a provided photo; additional information was requested from the customer, but no response was received.The complaint was not confirmed.The cause code was updated to reflect dc: traced to user - misuse/mishandling and defect code to needle - coring; according to the product specifications, syr100185 standard hypodermic needles are not designed to penetrate vials.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|