Spectranetics Corporation GlideLight Laser Sheath; Device, Removal, Pacemaker Electrode, Percutaneous
|
Back to Search Results |
|
| Model Number |
500-303 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Great Vessel Perforation (2152); Cardiac Perforation (2513); Pericardial Effusion (3271)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
A lead extraction procedure commenced to remove two right ventricular (rv) (left and right side) and two right atrial (ra) leads (left and right side) due to bacteremia.Spectranetics lld lead locking devices (llds) was inserted into each lead to provide traction.Beginning with a spectranetics 14f glidelight laser sheath on the left rv lead, adhesions in the vasculature were removed.Switching to a spectranetics 13f tightrail sub-c rotating dilator sheath, the left rv and left ra lead were removed successfully.Next, a spectranetics 16f glidelight laser sheath was used on both the right rv and right ra.When lasing on the right rv lead, the patient's blood pressure dropped and a pericardial effusion was detected via transesophageal echocardiogram (tee).Rescue efforts began, including rescue balloon, cpr, and sternotomy.A perforation at the inferior vena cava (ivc)/ra junction into the ra was discovered and repaired; however the patient did not survive procedure, and the right ra and rv lead remained in the patient.This report captures the 16f glidelight in use when the perforation occurred, requiring intervention, but resulting death.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
B7) other relevant history - not available from facility.D4) device serial number - not available from facility.H3) the glidelight was discarded, thus no product investigation was performed.H6) per ifu, perforation and death are listed as potential adverse events with use of glidelight devices.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Search Alerts/Recalls
|
|
|