The customer reported that small towel fibers or lint appear to be sticking to surgical gloves during procedures, and during an "ep procedure" on (b)(6) 2025, "a lot of blue towel fibers" were observed on gloves and the wire.The provided photo was reviewed, the provided sample was evaluated, the vendor was notified of the issue, and trend analysis was performed by reviewing the past two years of complaint data.The or towels were noted to shed lint, confirming the reported issue, and the product failed to meet specifications.The vendor noted that the cause of the issue is from lack of the vacuuming the product after the seaming process, and as a corrective action, vacuuming equipment will be added into the folding process; no further investigation into the cause and complaint is warranted, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|