| Model Number |
M00565020 |
| Medical Device Problem Codes |
Premature Activation (1484); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a wallflex esophageal stent was attempted to be implanted in the duodenum to treat a duodenal stricture during a stent placement for duodenal stricture procedure performed on (b)(6) 2026.It was reported that during procedure, the device would not fit down the pediatric colonoscope channel.Due to the tightness of the channel, the stent was inadvertently deployed in the channel.The stent was removed from the patient, and the procedure was completed using another of the same device.There is no patient complications reported as a result of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6:
imdrf device code a150103 captures the reportable event of stent premature deployment.
|
| |
|