| Model Number |
Harmony iQ 3600 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that the monitors connected to their harmonyiq 3600 integration system are flickering.No report of patient or procedure involvement.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and found that the fiber cable required replacement.The technician replaced the fiber cable, tested the function and operation of the system, confirmed it to be operating to specification, and returned the system to service.The unit was installed in 2017 making it approximately 9 years old.No additional issues have been reported.
|
| |
|