Model Number M635TU90050
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 09/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was being performed, and a watchman trusteer access system (was) and a watchman flx pro closure device and delivery system were selected to be used.During the procedure, the activated clotting time (act) was 368 seconds.Half of the heparin dose was administered prior to transseptal puncture, with the remaining half administered afterward.There were no partial or full recaptures performed, and the closure device was successfully deployed on the first attempt.An echogenicity was noted on the trusteer access system as the closure device was deployed.While removing the delivery system and trusteer sheath out, the physician aspirated.The delivery system was removed from the left side, and the echogenicity was not seen on left side.The scrub technician was able to flush the was sheath once it was out of the body and observed thrombotic material outside of the was sheath on distal tip.Per the physician, no thrombus was noted prior to transesophageal echocardiography (tee) removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN TruSteer? Access System
Common Device Name
Catheter, percutaneous
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50804100
Report Number2124215-2026-49685
Device Sequence Number16097441
Product Code DQY
UDI-Device Identifier00191506022310
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K240018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberM635TU90050
Device Catalogue NumberM635TU90050
Device Lot Number0039760636
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/17/2026
Date Device Manufactured07/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexMale
-
-