The account reported complaints that the suresite dressing vad1683 was not adhering properly and that it had peeled off of patients.A sample was received for investigation; 3 ea were received and 20 ea were pulled from inventory, and the samples returned and the samples from inventory had visual presence of the adhesive and passed functional testing by application and removal from a test surface.No defects were identified, and medline was unable to confirm the reported issue.At this time, the root cause cannot be determined for this issue, and this issue will be monitored by centurions complaint system.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|