It was reported that the patients output current was not being delivered, and lead impedance was >9,000 ohms, suggesting a possible lead discontinuity or hardware malfunction.Chest and neck x-rays were ordered to assess for lead damage or fracture.The patient reported recent attendance at a fair where ride restraints may have applied pressure over the implant site, potentially contributing to device- or lead-related trauma.Due to the diagnostic findings and concern for lead failure, surgical exploration with possible lead revision or replacement was scheduled.It was later noted that the x-rays were not available for review.The provider noted that the pin was fully inserted and that there was no visible lead break.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
|