| Model Number |
G5A-80A |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "hv charger test capacitor not holding charge" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly testing, stress testing, and functional testing without duplicating the report.The report was cleared and did not reoccur.An internal inspection of the device found no discrepancies.The main board was replaced to help prevent reoccurrence.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|