The customer reported that the safety clip breaks while in use.Medline received one case of safety blood collection needles in new condition, randomly selected 25 samples from each box for visual and functional testing, screwed the needles into medline branded blood collection tube holders, and activated the safety mechanism on each needle.The complaint was not confirmed; visual inspection revealed no manufacturing defects, no issues were experienced during functional testing, and there were no noticeable defects or issues noted when activating the safety mechanism.The cause was documented as not confirmed with received sample, and the samples worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|