Model Number 7840
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438); Off-Label Use (1494)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the health care professional (hcp) requested a non-conditional magnetic resonance imaging (mri) scan on device with high out of range pacing impedance measurements on this right atrial (ra) lead.Technical services (ts) was consulted for further review.Upon review, the stored electrogram (egm) showed a lead safety switch (lss) from bipolar to unipolar configuration due to high out of range pacing impedance measurements.Further review of stored atrial tachy response (atr) events showed ra lead also exhibited myopotential oversensing noise.It was suspected cause to be due to lead fracture.No adverse patient effects were reported.At this time, this ra lead remains in service.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY?+
Common Device Name
Implantable Lead
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50804608
Report Number2124215-2026-49695
Device Sequence Number10549007
Product Code NVN
UDI-Device Identifier00802526604430
UDI-Public(01)00802526604430(17)271111
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7840
Device Catalogue Number7840
Device Lot Number1160504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/17/2026
Date Device Manufactured11/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-