• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-MAJOR HEAD AND NECK PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-MAJOR HEAD AND NECK PACK Back to Search Results
Catalog Number DYNJ38985G
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that ¿the bovie pad and bovie are in the internal ¿sterilized¿ contents¿.It was later reported that ¿it was actually the bovie pad and the individual gown that the staff said was tucked inside the back table cover rather than on top of the pack as usual¿.
 
Additional Manufacturer Narrative
The account reported finding the pack assembled incorrectly, and an update received from the account stated it was actually the bovie pad and the individual gown that the staff said was tucked inside the back table cover rather than on top of the pack as usual.A sample was received, the account provided two sample packs, both sample packs were reviewed, the finished good production records were reviewed, and the finished good pack build was reviewed.Based on the complaint details and sample review, the gown component and grounding pad component were found to be correctly placed within the pack; therefore, the reported incorrect assembly was not confirmed.The cause code was dc: traced to user with knowledge, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-MAJOR HEAD AND NECK PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rahul kota
three lakes drive
northfield, IL 60093--275
MDR Report Key50804615
Report Number1417592-2026-43706
Device Sequence Number10549008
Product Code OGR
UDI-Device Identifier40195327385683
UDI-Public40195327385683
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ38985G
Device Lot Number24ABQ232
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/18/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-