The customer reported a leaking drainage bag and stated that urine was on the floor and the foley bag was leaking urine with a small hole in the lower left corner of the foley bag.Medline quality received 1 ea of the reported product in unpackaged, used condition; visual assessment revealed no defects nor abnormalities, and functional testing was conducted by filling the drainage bag with 800 ml of liquid and conducting a hang test to observe for leaks.No leaks were observed and could not be found at the reported location; the reported issue of drain bag leakage could not be verified based on visual or functional inspection of the received sample.The cause was not confirmed with received sample, and medline quality will continue to monitor for the issue through monthly complaint meetings and quarterly trending analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|