| Model Number |
IntelliVue Patient Monitor MX850 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the monitor did not alarm for an spo2 event due to a speaker malfunction.The device was in use on a patient at the time of the event.There was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by field service personnel.The logs indicated that the spo2 alarm was generated but was not heard by the user, and a speaker malfunction error was displayed.It was determined that the mainboard and speaker needed to be replaced.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a faulty mainboard and speaker assembly.The mainboard and speaker assembly were replaced to resolve the issue.The device was tested and performance was verified.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.Reporting institution phone # (b)(6).
|
| |
|