• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medizin Systeme Böblingen GmbH IntelliVue Patient Monitor MX850; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medizin Systeme Böblingen GmbH IntelliVue Patient Monitor MX850; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number IntelliVue Patient Monitor MX850
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the monitor did not alarm for an spo2 event due to a speaker malfunction.The device was in use on a patient at the time of the event.There was no adverse event reported.
 
Additional Manufacturer Narrative
Analysis was performed by field service personnel.The logs indicated that the spo2 alarm was generated but was not heard by the user, and a speaker malfunction error was displayed.It was determined that the mainboard and speaker needed to be replaced.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a faulty mainboard and speaker assembly.The mainboard and speaker assembly were replaced to resolve the issue.The device was tested and performance was verified.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.Reporting institution phone # (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IntelliVue Patient Monitor MX850
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
+49(703)14632031
MDR Report Key50804790
Report Number9610816-2026-101932
Device Sequence Number20993649
Product Code MHX
UDI-Device Identifier00884838070790
UDI-Public(01)00884838070790
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K210906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIntelliVue Patient Monitor MX850
Device Catalogue Number866470
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-