Ethicon Endo-Surgery, LLC. Harmonic ACE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH23
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the scalpel was hard to cut.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4) date sent: 9/17/2026 d4 batch#: a9ck1n investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was returned with the blade broken inside of the tube.During functional testing to the energy system, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device will stop activating when the blade becomes damaged.Due to the condition of the returned device, we were unable to test for the reported tissue effect issues.The device was disassembled to verify the blade and the blade breaking point was located at an area inside the tube proximal to the tissue pad.This is consistent with a side load being applied to the blade while active with the jaw closed.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.Continued usage of the damaged blade can result in a broken blade.No conclusion could be reached as to how the blade crack issue occurred.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch a9ck1n, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Harmonic ACE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50804911
Report Number3005075853-2026-07653
Device Sequence Number18794365
Product Code LFL
UDI-Device Identifier10705036014676
UDI-Public(01)10705036014676(11)280229(10)A9CK1N
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132612
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHARH23
Device Lot NumberA9CK1N
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/18/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/06/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1