investigation summary
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was returned with the blade broken inside of the tube.During functional testing to the energy system, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device will stop activating when the blade becomes damaged.Due to the condition of the returned device, we were unable to test for the reported tissue effect issues.The device was disassembled to verify the blade and the blade breaking point was located at an area inside the tube proximal to the tissue pad.This is consistent with a side load being applied to the blade while active with the jaw closed.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.Continued usage of the damaged blade can result in a broken blade.No conclusion could be reached as to how the blade crack issue occurred.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch a9ck1n, and no non-conformances were identified.
|