OGDEN MANUFACTURING PLANT OPTIFLUX F180NRe HF 12/CS 1.7SA EBEAM; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 0500318E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility¿s biomedical technician (bmt) reported that a visible internal dialyzer blood leak occurred five minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fme 2008t machine, did not alarm.Blood leak test strips were used and tested negative for the presence of blood.A fme bloodline was not in use.There was no obvious defect or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 250ml.The patient was restarted on a new machine and treatment completed successfully with new supplies.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F180NRe HF 12/CS 1.7SA EBEAM
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50804935
Report Number0001713747-2026-00350
Device Sequence Number8543059
Product Code KDI
UDI-Device Identifier00840861100156
UDI-Public00840861100156
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K162488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0500318E
Device Lot Number26EU07002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Supplement Date Received by Manufacturer09/28/2026
Initial Report FDA Received Date09/17/2026
Supplement Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FME Hemodialysis Machine
Patient SexFemale
Patient Weight176 LB
-
-