A customer reported their adc device is too hot to hold.There was no report of fire, smoke, explosion, or shock.There was no report of death, serious injury, or mistreatment associated with this event.
|
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Device history records (dhrs) for the libre reader was reviewed and the dhrs showed the libre reader passed all tests prior to release.At the time of complaint intake, this event was deemed not reportable and troubleshooting determined that no further investigation of the device was required.This complaint has been reassessed as reportable upon retrospective review associated with a capa and is being submitted with all available information.All pertinent information available to abbott diabetes care has been submitted.
|