| Model Number |
72400024 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
03/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurrent urinary incontinence.Imaging identified less or no fluid in the pressure regulating balloon; however, the source of the leak was unknown, as the device was not tested on the back table.A surgical procedure was performed in which the existing aus was explanted and replaced.There were no further patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant products implanted at the time of the event:
model number/catalog number: 72404132
serial number: n/a
batch/lot number: 1000446598
model/catalog description: belt cuff fgs 5.0 cm iz
unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127
serial number: n/a
batch/lot number: 1000414219
model/catalog description: control pump fgs iz
unique identifier (udi) #: (b)(4).This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (ncep-00181844) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.
|
| |
|