• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Ultraflex? Esophageal NG; prosthesis, esophageal

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Ultraflex? Esophageal NG; prosthesis, esophageal Back to Search Results
Model Number M00513730
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to boston scientific corporation that an ultraflex esophageal ng covered distal release stent was implanted in the esophagus to treat a 6 cm malignant esophageal stenosis during an endoscopic retrograde cholangiopancreatography procedure (ercp) performed on (b)(6) 2026.The patient's anatomy was tortuous and was dilated prior to stent placement.On (b)(6) 2026, post stent placement, it was noted under imaging examination that the stent had migrated.The stent was removed using biopsy forceps, and another of the same device was used to complete the procedure.There were no patient complications reported as a result of this event and the patient condition after the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a010402 captures the reportable event of stent migration.Imdrf impact code f2301 captures the reportable event of the migrated stent was removed.Imdrf impact code f2203 captures the reportable event of the imaging examination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ultraflex? Esophageal NG
Common Device Name
prosthesis, esophageal
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50805009
Report Number3005099803-2026-03835
Device Sequence Number8597422
Product Code ESW
UDI-Device Identifier08714729716150
UDI-Public(01)08714729716150(17)270219(10)35888650
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K130004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00513730
Device Catalogue NumberM00513730
Device Lot Number0035888650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/17/2026
Date Device Manufactured02/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient Weight132 LB
-
-