• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Penumbra, Inc. Velocity Delivery Microcatheter; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Penumbra, Inc. Velocity Delivery Microcatheter; DQY Back to Search Results
Catalog Number VEL160STR
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the right middle cerebral artery (mca) using a velocity delivery microcatheter (velocity), a penumbra system red 72 reperfusion catheter (red72), a neuron max 6f 088 long sheath (neuron max), and a guidewire.It was reported that the patient's anatomy was tortuous.The neuron max was placed in the right internal carotid artery (ica).While advancing the red72 and the velocity over the guidewire through the neuron max to the target location, resistance was encountered.The velocity and guidewire were removed together.After removal, the physician observed damage to the distal end of the velocity.While inspecting the velocity, the velocity was found to be fractured at the distal end.It was reported that the guidewire was still within the velocity.No part of the velocity remained in the patient.The physician completed the procedure with one pass using the red72.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Velocity Delivery Microcatheter
Common Device Name
DQY
Manufacturer (Section D)
Penumbra, Inc.
one penumbra place
alameda CA 94502
Manufacturer (Section G)
Penumbra, Inc.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica menezes
one penumbra place
alameda, CA 94502
+1(510)748-3200
MDR Report Key50805204
Report Number3005168196-2026-00320
Device Sequence Number15602626
Product Code DQY
UDI-Device Identifier00814548012629
UDI-Public00814548012629
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K100826
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVEL160STR
Device Lot NumberF00021753
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-