The customer advised pouches melted and failed the spore check for autoclave.Medline quality received 1 bx of the item for evaluation, and visual inspection was conducted with the provided sample; no abnormalities have been found.The complaint was determined to be a customer knowledge, as the customer is using the wrong sterilization pouch: they are currently using a tyvek pouch which is intended for low temperature sterilization, and they should be using a paper sterilization pouch which is intended for steam sterilization.The division will continue to monitor this issue for trending, and no credits or replacements will be issued at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|