| Model Number |
6935M62 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that post implant the right ventricular (rv) lead exhibited t-wave oversensing (twos) post ventricular pacing.The rv lead also triggered a lead integrity alert (lia) due to high rate non-sustained episodes and sensing integrity counter (sic).Repr ogramming was performed but twos was still present.Additionally, the right atrial (ra) lead exhibited far field r-wave (ffrw) oversensing.Reprogramming was also performed for the ra lead.The rv and ra leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|