• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Endo-Surgery, LLC. LIGAMAX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Endo-Surgery, LLC. LIGAMAX; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a lap distal panc.El5ml jammed.No delay or patient harm was reported.
 
Additional Manufacturer Narrative
(b)(4).Date sent 9/17/2026.D4 batch # z70540.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.To evaluate the reported incident, the device underwent functional testing.During testing, the instrument was fired, at which point the advancer tip was observed to be broken, preventing the clip from fully advancing into the jaws.To further assess the condition of the internal components, the device was disassembled for detailed inspection.Upon disassembly, two (2) clips were recovered from the clip track.The broken advancer tip was determined to have contributed to the observed clip-feeding issue.The event reported was confirmed and it is related to improper use of the device.Prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch z70540 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50805746
Report Number3005075853-2026-07657
Device Sequence Number8653586
Product Code FZP
UDI-Device Identifier10705036001843
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K110699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberZ70540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/31/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-