(b)(4).Date sent 9/17/2026.D4 batch # z70540.Investigation summary
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.To evaluate the reported incident, the device underwent functional testing.During testing, the instrument was fired, at which point the advancer tip was observed to be broken, preventing the clip from fully advancing into the jaws.To further assess the condition of the internal components, the device was disassembled for detailed inspection.Upon disassembly, two (2) clips were recovered from the clip track.The broken advancer tip was determined to have contributed to the observed clip-feeding issue.The event reported was confirmed and it is related to improper use of the device.Prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch z70540 number, and no non-conformances were identified.
|