(b)(4).Date sent 9/17/2026.D4 batch # a98l5k.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage, and the opened packaging was returned along with the instrument.To evaluate the reported incident, the device underwent functional testing.During testing, the instrument was fired; however, the clips failed to advance into the jaws.Further examination revealed that the tip of the advancer was bent, which prevented proper clip advancement.The device was subsequently disassembled for detailed internal inspection.Upon disassembly, the bent advancer was confirmed, and eight (8) clips were recovered from the clip track.The observed advancer deformation was determined to be consistent with the reported clip-feeding issue.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a98l5k number, and no non-conformances were identified.
|