• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-IV START KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-IV START KIT Back to Search Results
Catalog Number DYND74268
Medical Device Problem Codes Loss of or Failure to Bond (1068); Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the tegaderm is not sticking to the patient and when they go to peel the backing off it either does not come off or the backing comes off but the tegaderm stays stuck to it and the nurse is unable to peel it.".
 
Additional Manufacturer Narrative
The customer reported the tegaderm is not sticking to the patient and the backing either does not come off or the backing comes off but the tegaderm stays stuck to it and the nurse is unable to peel it.No sample or photo was received for the investigation; finished good production records and finished good pack build were reviewed, and there were no issues identified that likely contributed to the reported issue.The reported issue could not be confirmed without a sample.Due to other reports received, the likely root cause is a vendor issue relating to their manufacturing process, and medline is working with the component manufacturing facility for further evaluation and corrective actions.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-IV START KIT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
venkata peruri
three lakes drive
northfield, IL 60093--275
MDR Report Key50805808
Report Number1417592-2026-43881
Device Sequence Number20993657
Product Code LRS
UDI-Device Identifier40080196860440
UDI-Public40080196860440
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND74268
Device Lot Number24ABU677
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2024
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-