| Catalog Number |
142480489 |
| Medical Device Problem Code |
Free or Unrestricted Flow (2945)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A reported incident involving a primary plumset, pe lined tubing, 107.It was reported that the drip could not be locked and continued to flow.There was patient involvement; however, no patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone: (b)(6).One photo of the set in a bag was received for evaluation.The photo had been edited by the customer to place a red box around the cassette on the set.Review of the provided photo found no evidence of unrestricted flow.The used set was not returned for evaluation; therefore, the reported condition could not be replicated or confirmed.A device history review could not be completed due to the unknown lot number.
|
| |
|