It was reported that this pacemaker exhibited a code 1003 voltage alert, indicative that the voltage was too low for the projected remaining battery capacity.Technical services (ts) reviewed the available device data and noted that the device had recently been updated to newest software update.The voltage alert was determined to have likely triggered shortly after the software update due to activation of the monitoring algorithm.Engineering analysis confirmed that remote monitoring functionality remained available and that approximately six months of rf telemetry availability remained.Continued monitoring was recommended, until radio frequency (rf) telemetry is disabled, at which time replacement should be performed.No adverse patient effects were reported.This pacemaker remains in service.
|