• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Lumenis Ltd Lumenis Moses 200 D\F\L; powered laser surgical instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Lumenis Ltd Lumenis Moses 200 D\F\L; powered laser surgical instrument Back to Search Results
Model Number 1912-10
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, the fiber was found to be unusable.Another lot of the device was used to complete the procedure.There was no patient complication.After that, the fiber was returned for analysis and the microscope inspection found that the fiber connector had distorted light exiting the face, indicating an internal connector fracture.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this device was thoroughly analyzed.Visual inspection found that the fiber body was in good condition.Microscopic inspection found that the fiber connector had distorted light exiting the face, indicating an internal connector fracture.Additionally, the fiber tip was degraded.The connector showed burn marks on the face.During functional test, there was a weak aim beam exiting the fiber tip.Based on this information, the complaint was confirmed.Fracture within the connector can occur during procedural handling of the fiber during setup or use.This connector fracture can result in an insufficient aiming beam or low laser energy output, up to a complete inability for the fiber to fire.In addition, it is likely that the interaction between the fractured connector and blast shield led to the burn marks observed on the fiber face.A device history record (dhr) review indicates the device met all manufacturing specifications, and therefore the failure was unlikely related to manufacturing.The labeling review found that the product instructions for use (ifu) states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement and hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber damaged/defective is defined in the risk documentation.Based on this analysis, the investigation conclusion code assigned is cause traced to component failure which indicates expected or random component failure without any design or manufacturing issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Lumenis Moses 200 D\F\L
Common Device Name
powered laser surgical instrument
Manufacturer (Section D)
Lumenis Ltd
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
ADVANCED FIBER TOOLS GMBH
bornheimer str 4
mittweida 09468
GM   09468
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50806120
Report Number2124215-2026-49736
Device Sequence Number10589818
Product Code GEX
UDI-Device Identifier07290109143170
UDI-Public(01)07290109143170(17)280409(10)15762509
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1912-10
Device Catalogue Number1912-10
Device Lot Number0015762509
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/09/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-