MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic Back to Search Results
Model Number 72404127
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 08/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient implanted with this artificial urinary sphincter (aus) experienced recurrent incontinence approximately six months post-implantation and three months after device activation.The patient reported that the device was no longer functioning properly and that symptoms of incontinence had returned.During a clinical examination, it was determined that there was a potential device-related issue, and revision surgery was scheduled to investigate the cause.During the surgical procedure, after the pump was exposed, it was observed that saline was not returning from the pressure-regulating balloon to the pump as expected.Troubleshooting was performed; however, no cause could be identified.The saline volume within the balloon was measured and found to be 23 cc, consistent with the volume documented during the original implantation procedure.Based on this finding, it was initially concluded that the balloon was functioning properly.To minimize unnecessary component replacement, only the pump was explanted and replaced.The newly implanted pump was connected to the existing system, and the aus was tested through three complete cycles.The system functioned normally during all three cycles, suggesting that the issue was likely related to the original pump.Therefore, no additional components, including the cuff or pressure-regulating balloon, were explanted.No additional patient complications were reported.
 
Additional Manufacturer Narrative
Concomitant products implanted at the time of the event: model number/catalog number: 72404132.Serial number: n/a.Batch/lot number: 1100475240.Model/catalog description: belt cuff fgs 5.0 cm iz.Unique identifier (udi) #: ((b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100712529.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: ((b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50806220
Report Number2124215-2026-49739
Device Sequence Number8543099
Product Code EZY
UDI-Device Identifier00878953003078
Combination Product (Y/N)Y
Initial Reporter CountryIS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404127
Device Catalogue Number72404127
Device Lot Number1100719383
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/17/2026
Date Device Manufactured04/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexMale
-
-