(b)(4).Date sent: 9/17/2026.D4: batch #118e11.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the long60a device was received for analysis with no apparent damaged and with two gst60g reloads (b, c) present.Both reloads were received partially fired 1/5.The device was tested for functionality in the articulated position with the returned reloads by resetting and reloading them into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple lines and cut lines were complete and the remaining staples meet the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.As part of ethicon ¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 118e11, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished long60a, with lot number a9mc6k, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|