Model Number BF-UC180F
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the videoscope to the service center for evaluation related to a separate issue.During the evaluation, a reportable malfunction was identified: the acoustic lens was chipped.As a result, the affected component required replacement.There was no patient involvement associated with this event.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection based on the results of the investigation, the probable cause of this complaint is expected or random component failure resulting from physical stress, with no design or manufacturing issue identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Common Device Name
ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50806600
Report Number3002808148-2026-32741
Device Sequence Number16096061
Product Code PSV
UDI-Device Identifier04953170356353
UDI-Public04953170356353
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K070983
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-UC180F
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/18/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-