A complaint was received related to a strikethrough, and the customer reported "fluid soaked through" while "cleaning trays" while leaning against the sink.The samples were inspected and the w.L.Gore elbow lean test was performed on the body and sleeve of the gown; in both tests there was no strikethrough the material, and a review of the contract manufacturer¿s records, manufacturing process, and retained samples found no abnormalities or defects in their production.Based on the results of the investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.No root cause was identified in the record.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|