• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; GOWN,LVL4,OPENBK,THUMBLOOP,OVRHD,BLU,REG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; GOWN,LVL4,OPENBK,THUMBLOOP,OVRHD,BLU,REG Back to Search Results
Catalog Number NONLV410
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that lourdes spd tried this gown and found that it leaked fluid on the arms and abdomen.
 
Additional Manufacturer Narrative
A complaint was received related to a strikethrough, and the customer reported "fluid soaked through" while "cleaning trays" while leaning against the sink.The samples were inspected and the w.L.Gore elbow lean test was performed on the body and sleeve of the gown; in both tests there was no strikethrough the material, and a review of the contract manufacturer¿s records, manufacturing process, and retained samples found no abnormalities or defects in their production.Based on the results of the investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.No root cause was identified in the record.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
GOWN,LVL4,OPENBK,THUMBLOOP,OVRHD,BLU,REG
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
pranay abbu
three lakes drive
northfield, IL 60093--275
MDR Report Key50806732
Report Number1417592-2026-43945
Device Sequence Number12924916
Product Code FYA
UDI-Device Identifier10197344039586
UDI-Public10197344039586
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K213177
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberNONLV410
Device Lot Number49825090001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/20/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Initial Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-