• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Endo-Surgery, LLC. ECHELON Flex ENDOPATH; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Endo-Surgery, LLC. ECHELON Flex ENDOPATH; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PLE60A
Medical Device Problem Codes Failure to Cut (2587); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a gastroplasty procedure, it was observed that the device had an unspecified malfunction.There was no patient consequence nor surgical delay reported.Additional information requested.
 
Additional Manufacturer Narrative
(b)(4) date sent: 9/17/2026 d4: batch #unk h4; unknown.Additional information was requested, and the following was obtained: we hereby inform you of the occurrence of a technical failure involving surgical staplers used during colectomy and gastroplasty procedures, performed at the surgical center of the.During the procedures, the device was activated with the purpose of stapling and sectioning the tissue.After activation, it was observed that the stapler closed the staples, but did not section the tissue, remaining locked.In view of the impossibility of opening the device by the usual mechanism, an attempt was made to open it manually, without success.It was then necessary to use surgical forceps to promote the forced opening of the stapler, allowing its removal.Due to the intercurrence and the impossibility of using the device initially used, it was necessary to open a new surgical stapler to continue the procedure.In view of what happened, we request the technical evaluation of the device, as well as the position of this company regarding the occurrence and the appropriate measures, including the analysis of the unit involved and the verification of possible recurrence of the problem in other devices of the same batch.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: per aei received it states "we hereby inform you of the occurrence of a technical failure involving surgical staplers used during colectomy and gastroplasty procedures, performed at the surgical center of the (please refer to the attached mail).During the procedures".Please clarify the following: can you please confirm whether the previous events mentioned were reported to ethicon? if yes, could you please provide the corresponding product complaint (pc) number(s)? did those prior events involve the same device? approximately how many times has the surgeon observed this same issue? did any of the prior events result in patient impact or require a change to the procedure please confirm lot number this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON Flex ENDOPATH
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50807215
Report Number3005075853-2026-07668
Device Sequence Number15616542
Product Code GDW
UDI-Device Identifier10705036014300
UDI-Public(01)10705036014300(17)260531(10)A9D90C
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue NumberPLE60A
Device Lot NumberA9D90C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Supplement Date Received by Manufacturer09/29/2026
Initial Report FDA Received Date09/17/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-