It was reported that during the surgical procedure, two 1.1 mm drill bits belonging to the hand 1.5 module broke.The incident occurred because the supplies did not have the necessary edge, which forced the specialist to apply excessive force during drilling.Faced with this situation, the specialist expressed his disagreement and requested the immediate replacement of both worn drill bits.It also required the replacement of the t4 screwdriver bit, which is rotated, a condition that makes it difficult to correctly fasten and handle the screws.There was no patient harm.
|
|
Product complaint # (b)(4)
this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative:
d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6:
a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the lot code provided is not a valid finished goods lot.This complaint involves one sd scrwdrvr shft/t4 50/self-retain/hxc device.The product was not returned to depuy synthes; however, photos were provided for review.The photo investigation revealed that "03.130.010
sd scrwdrvr shft/t4 50/self-retain/hxc" has twisted.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces such as drilling the pin into the hole in an misaligned position.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the sd scrwdrvr shft/t4 50/self-retain/hxc would contribute to the complained device issue.Based on the investigation findings, the ¿screw driver¿ has ¿twisted¿ because of unintended force, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|