| Model Number |
60005 |
| Medical Device Problem Code |
Optical Decentration (1360)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A site reported to rxsight that a patient presented to the clinic with complaints of shadowing.During postoperative examination, the light adjustable lens (lal) was observed to be decentered.It was subsequently noted that one haptic was positioned in the sulcus rather than within the capsular bag.A secondary surgical procedure was performed, during which the lens was successfully repositioned.No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.Manufacturer reference #: (b)(4).
|
| |
|