Abiomed, Inc. - 1220648 Impella; Temporary Non-Roller Type Left Heart Support Blood Pump
|
Back to Search Results |
|
| Catalog Number |
0052-3046 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemolysis (1886); Ischemia (1942)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Impella cp smartassist was inserted via right femoral artery access site in a 55-year-old male patient presenting with acute myocardial infarction and cardiogenic shock.The patient¿s comorbidities were peripheral arterial disease.The patient¿s clinical state prior to initiation of support was society for cardiovascular angiography and interventions (scai) shock stage b.Stage b patients have clinical evidence of hypoperfusion that initially requires pharmacological or mechanical support.Hypotension is usually present.Prior to the impella placement the patient was on respiratory support.Following femoral insertion of the impella cp smartassist, the medical team observed signs of limb ischemia in the accessed extremity, including a cold foot.Due to concerns for ischemia, the impella cp smartassist was explanted after approximately 12 hours of support and therapy was escalated to an axillary artery inserted impella 5.5 smartassist.During support with the impella cp smartassist, the patient was also noted to have signs consistent with hemolysis.Urine discoloration was observed and plasma free hemoglobin levels were reported to be elevated.After transition to the impella 5.5 smartassist, plasma free hemoglobin remained elevated at 188 mg/dl on the following day but subsequently trended downward to 27 mg/dl and 13 mg/dl on later assessments.Prior to explant, the impella cp smartassist functioned as intended at performance level p-6 with reported flow of 3.0 l/min utilizing 5% dextrose in water with sodium bicarbonate purge solution.The patient remained on impella 5.5 smartassist support at the time of reporting.Limb ischemia is a known procedural complication associated with arterial access.Contributing factors may include the sheath diameter relative to vessel size, known peripheral arterial disease, vasoconstrictor use, and potentially subtherapeutic anticoagulation.Anticoagulation parameters, including activated clotting time (act), were not documented for review.The reported hemolysis is likely multifactorial and may be associated with mechanical circulatory support, which is known to increase red blood cell shear stress.
|
| |
|
Event or Problem Description
|
|
Clinical narrative updated 2026-9-23:
since original reporting, the patient outcome has been updated.The impella cp smartassist was explanted and replaced with an impella 5.5 smartassist.Despite continued support with the impella 5.5 smartassist, along with inotropic therapy, vasopressors, and respiratory support, the patient's clinical condition remained consistent with society for cardiovascular angiography and interventions (scai) stage d cardiogenic shock.After 7 days of support with the impella 5.5 smartassist, a decision was made to withdraw care and the patient expired.The previously reported limb ischemia and hemolysis may have contributed to the patient's overall clinical deterioration.Given the severity of the patient's underlying cardiogenic shock, the need for multiple escalated mechanical circulatory supports, and the occurrence of these complications, the patient's death is most likely attributable to the progression of the underlying critical illness.However, based on the information available, the contribution of impella support and the reported complications to the patient's outcome cannot be excluded.Therefore, the pump use cannot be dissociated from the patient outcome.Prior clinical narrative:
impella cp smartassist was inserted via right femoral artery access site in a 55-year-old male patient presenting with acute myocardial infarction and cardiogenic shock.The patient¿s comorbidities were peripheral arterial disease.The patient¿s clinical state prior to initiation of support was society for cardiovascular angiography and interventions (scai) shock stage b.Stage b patients have clinical evidence of hypoperfusion that initially requires pharmacological or mechanical support.Hypotension is usually present.Prior to the impella placement the patient was on respiratory support.Following femoral insertion of the impella cp smartassist, the medical team observed signs of limb ischemia in the accessed extremity, including a cold foot.Due to concerns for ischemia, the impella cp smartassist was explanted after approximately 12 hours of support and therapy was escalated to an axillary artery inserted impella 5.5 smartassist.During support with the impella cp smartassist, the patient was also noted to have signs consistent with hemolysis.Urine discoloration was observed and plasma free hemoglobin levels were reported to be elevated.After transition to the impella 5.5 smartassist, plasma free hemoglobin remained elevated at 188 mg/dl on the following day but subsequently trended downward to 27 mg/dl and 13 mg/dl on later assessments.Prior to explant, the impella cp smartassist functioned as intended at performance level p-6 with reported flow of 3.0 l/min utilizing 5% dextrose in water with sodium bicarbonate purge solution.The patient remained on impella 5.5 smartassist support at the time of reporting.Limb ischemia is a known procedural complication associated with arterial access.Contributing factors may include the sheath diameter relative to vessel size, known peripheral arterial disease, vasoconstrictor use, and potentially subtherapeutic anticoagulation.Anticoagulation parameters, including activated clotting time (act), were not documented for review.The reported hemolysis is likely multifactorial and may be associated with mechanical circulatory support, which is known to increase red blood cell shear stress.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc or its employees that the report constitutes an admission that the product, abiomed inc, or its employees, caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 clinical narrative updated and h6 codes were updated.H1 type of reportable event was updated from serious injury to death.
|
| |
|