| Model Number |
CDS0801-XTW |
| Medical Device Problem Code |
Difficult or Delayed Positioning (1157)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 4 degenerative mitral regurgitation (mr) for a mitraclip procedure.During the procedure, mitral valve height was 3.25 to 3.5 centimeters.Physician decided to proceed with this height.Once the clip was in position, physician unstraddled the guide to gain the height.This move has caused tear in the anterior leaflet.Intraaortic balloon pump was inserted but was unable to insert due to the small size of the vessel.Procedure was discontinued.All steps were performed as per ifu.Imaging was difficult.The final mr was grade 4.On (b)(6) 2026, it was reported that patient passed away.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|