| Model Number |
G447 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the medical doctor (md) called technical services (ts) questioning the right ventricular (rv) sensing and pacing for this cardiac resynchronization therapy defibrillator (crt-d).Ts reviewed the data and noted the presenting electrogram (egm) showed rv pace followed by atrial sensing and that sometimes there was likely fusion for two beats.Ts also noted there was a pacemaker mediated tachycardia (pmt) episode in late august and noise on the rv rate sense channel that was not sensed.Ts also discussed there was one beat of right atrial (ra) oversensing and an episode with a premature ventricular contraction (pvc).Ts stated there were no anomalies with the rv sensing that they noticed.Information from the field was received and noted the patient would continue to be monitored.No adverse patient effects were reported.This crt-d remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|