The customer reported the secondary was backflowing into the primary bag, and returned the secondary bag along with the primary suspect, the primary bag.The primary suspect is material 2426-0007, lot unknown.Upon initial visual examination, the sets had no defects or abnormalities.The sets were set up and hung with the primary on a hanger, and was run at 180 vtbi for 30 minutes on an alaris pump.The secondary set was infusing blue dyed water, and the primary tap water, in order to visualize the backflow.Even at very high flow rates, the back check valve held, and did not allow for any blue water past it towards the primary bag.The complaint of backflow could not be verified.A device history record review was not performed as the lot number is "unknown" for this complaint.The root cause of this issue could not be identified without a replicated failure.
|