It was reported that during a device upgrade procedure, defibrillation threshold (dft) testing was performed but was unsuccessful as the right ventricular (rv) defibrillation lead could not detect and convert.The paced qrs was 2.5mv, the induced vf was not detected and an external shock was delivered to convert the arrhythmia.The rv lead was repositioned, sensitivity was adjusted and the paced qrs was now 4.2mv.Dfts failed at 21 joules and 31 joules in the initial polarity, then was successful at 31 joules in reversed polarity.In total, 11 dfts were performed.The patient passed away over the weekend; there were no allegations against the implanted device but the physician was looking to see if any episode egms were available.Technical services discussed there was no latitude data and recommended interrogating the device post-mortem.It was noted that the patient had undergone a coronary artery bypass graft (cabg) one week before.Additional information was received confirming the device was not interrogated post-mortem and the device and lead were not explanted for return.The official cause of death was not provided to the local sales representative and the physician did not consider the device to be a contributory cause in the death.However, boston scientific is not able to rule out the prolonged procedure with multiple dfts as a contributing cause.
|